仪器分类
高级光合荧光测量仪
高级光合荧光测量仪
仪器编号
2024110100015
SN序列号
SN:68C-373469
生产厂家
美国
型号
LI-6800
制造国家
美国
分类号
放置地点
ES BuildingES432
购置日期
2024-11-30
入网日期
2025-04-01

主要规格及技术指标

Analyzer:

1.1 Analyzer Position: The infrared analyzer must be located at the head of the leaf chamber to achieve synchronization between the reference chamber and sample chamber measurements.
1.2 CO2 Range: 0-3000 µmol mol⁻¹
1.3 CO2 Accuracy: At 400 μmol/mol, signal noise RMS ≤ 0.1 μmol/mol @ 4s signal.
1.4 H2O Range: 0-75 mmol mol⁻¹
1.5 H2O Accuracy: At 10 mmol/mol, signal noise RMS ≤ 0.01 mmol/mol @ 4s signal.
Gas Flow Rate: Not less than 1000 µmol s⁻¹.

Main Unit Pressure Sensor:

3.1 Measurement Range: 50~110 kPa
3.2 Accuracy: ±0.4 kPa
3.3 Resolution: 1.5 Pa
Leaf Chamber Pressure Sensor:

4.1 Pressure Control Range: 0~0.1 kPa
4.2 Resolution: <1 Pa
4.3 Signal Noise: Average 1 Pa @ 4s signal
Quantum Light Sensor:

5.1 Detector: Silicon photodetector
5.2 Sensitivity: 5~10 µA / 1000 µmol m⁻² s⁻¹
Temperature:

6.1 Operating Temperature: 0~50℃
6.2 Storage Temperature: -20℃~60℃
6.3 Leaf Chamber Temperature Control Range: ±10% of ambient temperature
Fluorescence:

7.1 Modulation Frequency: 1 Hz~250 kHz
7.2 Activation Light: 0~3000 µmol m⁻² s⁻¹
7.3 Saturation Light: Not less than 12000 µmol m⁻² s⁻¹
7.4 The fluorescence leaf chamber can simultaneously measure photosynthesis and rapid fluorescence induction kinetics curves (O-J-I-P).
7.5 Fluorescence Signal Temperature Dependence: ±0.25% / ℃
Main Unit:

8.1 Storage: Not less than 2G
8.2 Display: Touchscreen operation, resolution 1024×600, display can rotate freely 180 degrees.
8.3 Processor: ARM® Cortex™ A9 quad-core, 1GHz
8.4 Humidification Tube: Isolated with an ion exchange membrane, pure water used to control humidity.
8.5 Equipped with Tripod: Can secure both the main unit and the leaf chamber.
1.分析器: ★1.1分析器位置:红外分析器必须位于叶室头部,以实现参比室和样品室测量的同步性 1.2 CO2量程:0-3000 µmol mol-1 1.3 CO2精度:400 μmol/mol时,信号噪声RMS ≤0.1 μmol/mol@4s信号 1.4 H2O量程:0-75mmol mol-1 1.5 H2O精度:10 mmol/mol时,信号噪声RMS ≤0.01 mmol/mol@4s信号 ★2.气体流速:最高不低于1000 µmol s-1 3.主机压力传感器: 3.1测量范围:50~110 kPa 3.2准确度:±0.4 kPa 3.3分辨率:1.5 Pa 4.叶室压力传感器: ★4.1控压范围:0~0.1 kPa 4.2分辨率:<1 Pa 4.3信号噪声:平均1 Pa@4s信号 5.光量子传感器: 5.1探测器:硅光电探测器 5.2灵敏度:5~10 µA /1000 µmol m-2 s-1 6.温度: 6.1工作温度:0~50℃ 6.2存储温度:-20℃~60℃ 6.3叶室控温范围:环境温度的±10℃ 7.荧光: 7.1调制频率:1 Hz~250 kHz 7.2活化光:0~3000 µmol m-2s-1 ★7.3饱和光:不低于12000 µmol m-2s-1 ★7.4荧光叶室可实现光合和快速荧光诱导动力学曲线(O-J-I-P)的同时测量 7.5荧光信号温度依赖性:±0.25% / ℃ 8.主机: 8.1存储:不低于2G 8.2显示:可触屏操作,分辨率1024×600,显示屏可180度自由旋转 8.3处理器:ARM®CortexTM A9四核,1GHz 8.4加湿管:内有离子交换膜隔离,加纯水控制湿度 8.5配备三脚架:既可固定主机,也可固定叶室

主要功能及特色

Gas exchange measurement is an effective method for studying the photosynthetic capacity at the leaf scale of plants. This instrument measures gas exchange parameters of horticultural plants, crops, and other plant leaves, such as photosynthetic rate, transpiration rate, and chlorophyll fluorescence parameters. This research helps to understand the physiological and ecological characteristics of plants, providing scientific guidance and strong support for the analysis of photosynthetic physiological traits, trait analysis, and the selection of superior crop varieties, as well as the study of environmental tolerance mechanisms.

气体交换测量是研究植物叶片尺度的光合能力的有效方法。该仪器通过测量园艺植物、农作物等植物叶片的气体交换参数,如光合速率、蒸腾速率、叶绿素荧光参数等指标,以此研究植物的生理生态特性,对植物的光合生理特征、性状分析以及优良作物品种筛选、耐受环境机制研究起到科学的指导作用和强有力的科研支撑。